Installation of Star 430 series turbines:
- REMOVE THE END CAP
- CLEAN OUT THE HANDPIECE HEAD USING A COTTON SWAB, MAKE SURE ALL DEBRIS IS REMOVED
- CLEAN OUT THE END CAP USING A COTTON SWAB MAKING SURE ALL DEBRIS IS REMOVED ( YOU CAN USE ALCOHOL, LUBRICANT, OR HANDPIECE CLEANER TO HELP WITH THE CLEANING PROCESS )
- PUSH THE TURBINE INTO THE END CAP MAKING SURE THE REAR GREEN O-RING IS SEATED COMPLETELY AND PROPERLY. YOU SHOULD NO LONG BE ABLE TO SEE THE GREEN O-RING AND THE END CAP SHOULD STAY ON THE REAR BEARING BY ITSELF.
- HOLDING THE HANDPIECE WITH THE OPENING FOR THE TURBINE POINTING DOWN, INSERT THE TURBINE UP INTO THE HEAD AND SCREW ON THE END CAP.
- CONNECT THE HANDPIECE TO YOUR CONTROL UNIT AND RUN FOR 30 SECONDS. THE TURBINE MAY RUN SLOW AT FIRST AS THE GREASE IS DISTRIBUTED THROUGHOUT THE BEARINGS.
| |
| | |
|
| | |
|