Installation of Star 430 series turbines:
  1. REMOVE THE END CAP
  2. CLEAN OUT THE HANDPIECE HEAD USING A COTTON SWAB, MAKE  SURE ALL DEBRIS IS REMOVED
  3. CLEAN OUT THE END CAP USING A COTTON SWAB MAKING SURE ALL DEBRIS IS REMOVED ( YOU CAN USE ALCOHOL, LUBRICANT, OR HANDPIECE CLEANER TO HELP WITH THE CLEANING PROCESS )
  4. PUSH THE TURBINE INTO THE END CAP MAKING SURE THE REAR GREEN O-RING IS SEATED COMPLETELY AND PROPERLY.  YOU SHOULD NO LONG BE ABLE TO SEE THE GREEN O-RING AND THE END CAP SHOULD STAY ON THE REAR BEARING BY ITSELF.
  5. HOLDING THE HANDPIECE WITH THE OPENING FOR THE TURBINE POINTING DOWN, INSERT THE TURBINE UP INTO THE HEAD AND SCREW ON THE END CAP.
  6. CONNECT THE HANDPIECE TO YOUR CONTROL UNIT AND RUN FOR 30 SECONDS. THE TURBINE MAY RUN SLOW AT FIRST AS THE GREASE IS DISTRIBUTED THROUGHOUT THE BEARINGS.